IJPRS

Home Article Development and Validation of pH-Independent Spectroscopic Method for Estimation of Gemcitabine HCl in Pharmaceutical Formulation


Research Article

Development and Validation of pH-Independent Spectroscopic Method for Estimation of Gemcitabine HCl in Pharmaceutical Formulation


Author(s)

Patel, P., Patel P.U.


Author's Affiliation


Abstract

This method has been developed by measuring the absorbance of Gemcitabine HCL at 254 nm (pH 3.0 – 10.9).   The method is simple, accurate and precise, giving linearity in the range of 14 – 34 μg/mL for Gemcitabine HCL with R2= 0.999 (n= 6).  The method was employed in testing the concentration of Gemcitabine HCL in marketed formulations, and the results were found in agreement with the labeled amount.


Keywords

UV spectrophotometry, Isosbestic point, pH independent, Gemcitabine HCL


Cite This Article

Patel, P., Patel P.U. (2017). Development and Validation of pH-Independent Spectroscopic Method for Estimation of Gemcitabine HCl in Pharmaceutical Formulation. , 6(2), 140 - 145.


Leave a Reply

Your email address will not be published. Required fields are marked *

Recent Articles